MA-1 Smart Karl Fischer Moisture Analyzer Installation Site Description

MA-1 Smart Karl Fischer Moisture Analyzer Installation Site Description Installation and location of a moisture meter:
1. The Karl Fischer moisture analyzer should not be installed in a room with corrosive gas, which will partially corrode the instrument circuit and shorten the life of the instrument.
2. The room temperature is below 5 degrees or above 40 degrees.
3. Do not install in places with high sunlight and humidity.
4. Do not install in an unstable power supply. It is recommended to use an electronic voltage regulator.

Maintenance of the two electrodes:
1. When the measuring electrode is contaminated, it can be wiped with acetone or absolute ethanol. If it is still not decontaminated, the platinum ball can be burned with an alcohol lamp.
2. When the electrolysis electrode is contaminated, acetone or absolute ethanol can be injected into the cathode chamber and shaken repeatedly to remove internal dirt.

Drying of the three electrolytic cells:
The electrolytic cell was decomposed and dried in an oven of about 50 degrees for five hours, and the reagent was added after it was naturally cooled.

MA-1 intelligent Karl Fischer moisture analyzer technical parameters
1. Measurement range: 10ppm-100% (H2O mass fraction)
2. Moisture content resolution: 1ppm
3. Measuring tube resolution: 0.001
4. Water titration repeatability: ≤0.005
5. Linear correlation coefficient of water titration: ≥0.999
6. Capacity error ≤ ± 0.0002
7. Relative error ≤ 0.2%